Installation and operation method of 2X series two-stage rotary vane vacuum pump

The newly purchased 2X series two-stage rotary vane vacuum pump must carefully read the installation method of the rotary vane vacuum pump and the operation method of the rotary vane vacuum pump before installation, otherwise the vacuum pump may be easily damaged.
Rotary vane vacuum pump
The newly purchased 2X series two-stage rotary vane vacuum pump must carefully read the installation method of the rotary vane vacuum pump and the operation method of the rotary vane vacuum pump before installation, otherwise the vacuum pump may be easily damaged.

The operation method of 2X series two-stage rotary vane vacuum pump has the following points;

1. The rotary vane vacuum pump should be installed in a horizontal position and cannot be tilted or inverted. The space of the rotary vane vacuum pump should be kept at least 300 mm to ensure that the rotary vane vacuum pump can dissipate heat well. The space of the oil level should be sufficient to facilitate normal operation. People can check the oil level frequently.

2. When the rotary vane vacuum pump is running, the direction of rotation of the motor is the same as the arrow mark on the vacuum pump.

3. Under normal circumstances, the 2X series double-stage rotary vane vacuum pump manufactured by the factory has not been refueled. Before the first use, the vacuum pump oil of No. 100 should be injected into the oil chamber through the fuel filler port. The oil level is added to the glass from the outside of the glass mirror. The position of /4 can be used. (The oil level of the single-stage rotary vane vacuum pump should not be lower than 1/2 of the height of the glass mirror).

4. The inlet of the two-stage rotary vane vacuum pump is only equipped with a metal filter. If the gas to be pumped contains dust or particles, a suitable air inlet filter should be installed at the inlet of the rotary vane vacuum pump. When the gas to be pumped contains water or contains corrosive gas, an adaptive gas-liquid separation device should be installed before the inlet of the single-stage rotary vane vacuum pump, otherwise it will affect the normal operation of the single-stage rotary vane vacuum pump.

5. The long-running rotary vane vacuum pump must be equipped with an automatic oil return pipe. The stored oil of the exhaust port is adjusted and controlled by the switch of the return pipe to control the oil return and the degree of vacuum.

Vacuum pump related products are: direct-connected vacuum pump, laboratory vacuum pump
Http://NewsDetail326T2.html

China Hardware Business Network http://news.chinawj.com.cn Editor: (Hardware Business Network Information Center) http://news.chinawj.com.cn

Touch Kitchen Faucets

Touch Kitchen Faucets,Touch Kitchen Faucets Price,Touch Kitchen Faucets Product

Yuyao Zelin Sanitary Ware Co., Ltd , https://www.zelinsinkfaucet.com